Installing Dionex Devices
Dionex ESA Model 582/584 HPLC Pumps: Installation
Device Type: |
Pump for HPLC applications |
Device Driver: |
Dionex HPLC - ESA Modules: ESA Pump The driver was tested with ESA pump model 584, firmware version 584 V 1.01. |
Restrictions: |
No flow ramps, only step-like changes of flow possible |
What is required?
License: |
Timebase Class 1 |
Connection: |
|
To control the instrument, use |
Hardware Installation
Device Connection: |
To establish communication between the pump and Chromeleon connect the RS-232 and the Fiber Optic cable as follows: Connection of the Fiber Optic Cable:
Connection of the RS-232 cable:
Note: The pump cannot be controlled via a UCI interface. |
Device Settings: |
The communication parameters are set at the factory as follows: Baud rate: 9600 Databits: 8 Stopbits: 1 Parity: EVEN Handshake: None Verify that the pump is correctly installed and configured in the Server Configuration program and that the instrument itself is switched to remote operation:
All commands are now executed via Chromeleon. Tip: For detailed information on device connections, please also refer to the Operating Manual of the Model 582 or 584 Solvent Delivery Module. |
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
On each configuration tab page, select the desired settings. For information about the settings on a page, click Help or press F1. On the ESA Pump Configuration page, in the COM Port list, select the serial interface port to which the pump is connected (i.e., the COM port on the Chromeleon computer to which the RS-232 serial cable is connected).
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For an overview of the different Dionex devices, refer to Installing Dionex Devices.